Our economical cold lasers allow you to offer affordable low level laser therapy to your patients increasing patient satisfaction, increasing referrals and your income.
Many of our clients report great success using treatment times of about six minutes per area with the 635nm 5mw wand or if using the 808nm 150mw wand about two minutes per treatment area. The 5 diode 635nm 808nm laser is usually used about 90 seconds per area. The lasers should be kept in continuous motion around the area(s) being treated.
Digital Hand Held Laser Controller Suggested Set Up:
Suggested set up:
Put laser goggles on everyone in the room
Plug in laser wand taking care to align the pins when inserting
(insert and remove using bullet connector not laser tip).
Insert and turn key – LCD will light up.
Hold “Set” till “Hz” flashes. (try 12Hz to start) Adjust Hz rate with Up/Down Buttons
Press “Set” – “Sec” flashes. (try 90 seconds to start) Adjust Seconds with Up/Down Buttons
Press “Set” to select.
Press RED Button to start or pause
Sec. counts down & green LED flashes during operation.
Use DC socket to charge battery or power if not charged.
LCD bat. charge lines are proportional to battery charge.
As long as the key is on the battery will be depleted.
Therefore turn the key off and remove it when the unit is not in use.
LIGHT FROM 808 NM LASERS IS NEARLY INVISIBLE.
TO DETERMINE IF 808NM LASERS ARE WORKING POINT THE LASER AT THE RAISED RECTANGLE ON THE
ENCLOSED YELLOW IR DETECTOR CARD (YOU SHOULD SEE AN ORANGE GLOW) OR AT WHITE SHEET OF PAPER IN A DARK ROOM, (YOU SHOULD SEE A SMALL RED DOT).
DO NOT LOOK INTO THE BEAM BECAUSE IT IS HARMFUL!
Equipment and components are distributed by and for Healing Lasers Design Company, Cape Girardeau, MO
FDA Facility Establishment Registration Number: 3004857915 and are sold by Accurate Inspections, Inc., a New Jersey Corporation doing business as Laser-Therapy.US.
Desktop Suggested Set Up: (sugested set up for portable & multi laser wand are listed below)
1. Put laser goggles on everyone in the room
2. Turn on power switch on back of unit. Slide the wand into the connector on top left corner of the power supply and twist 1/4 turn to the right to lock.
3. Push the FIRST button TWO times. The LEFT display should read 2.
4. Push the SECOND button TWO times for (2 Hz +/- to 20 Hz +/-). The RIGHT display should read 2
OR THREE times for (20 Hz +/- to 200 Hz +/-). The RIGHT display should read 3
5. Push the THIRD (RUN) button ONCE. The unit is now set to run in the selected range with a square wave output to the wands.
6. The VOLTAGE knob is the lower right knob. It controls voltage to the laser wand.
Voltage is displayed on the right display. Set the wand operating voltage to 9.5 volts.
7. Adjust the pulse rate of the laser light flashing with the PULSE RATE
(lower left knob). The pulse rate is displayed on the left display.
* The VOLTAGE knob is the lower right knob. It controls voltage to the laser wand.
Voltage is displayed on the right display. Set the voltage to 9.5 volts.
* The DUTY CYCLE knob is usually set to the red line for a crisp pulse. This upper right knob determines how long the pulse width is to the wand. We recommend 75% power on (duty) setting.
* The STAND-BY buttons are both left outward when the laser is in use. If either or both buttons are pushed inwards, voltage is lowered and the wands will NOT light.
The most common reason the wand will not light up is one or both of the STAND-BY buttons have been pushed, reducing the voltage.
LIGHT FROM THE 808 NM WAND IS NEARLY INVISIBLE. TO DETERMINE IF IT IS WORKING POINT THE WAND AT A WHITE SHEET OF PAPER IN A DARK ROOM.
NOT LOOK INTO THE BEAM BECAUSE IT IS HARMFUL!
Equipment and components are distributed by and for Healing Lasers Design Company, Cape Girardeau, MO
FDA Facility Establishment Registration Number: 3004857915
and are sold by Accurate Inspections, Inc., a New Jersey Corporation doing business as Laser-Therapy.US.